Circulation. 1999;100:826-831

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yamanouchi, Y.
Right arrow Articles by Tchou, P. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yamanouchi, Y.
Right arrow Articles by Tchou, P. J.
Related Collections
Right arrow CPR and emergency cardiac care
Right arrow Arrhythmias, clinical electrophysiology, drugs

(Circulation. 1999;100:826-831.)
© 1999 American Heart Association, Inc.


Clinical Investigation and Reports

Fully Discharging Phases

A New Approach to Biphasic Waveforms for External Defibrillation

Yoshio Yamanouchi, MD; James E. Brewer, MS; Kenneth F. Olson, BS; Kent A. Mowrey, MS; Todor N. Mazgalev, PhD; Bruce L. Wilkoff, MD; Patrick J. Tchou, MD

From the Department of Cardiology, Cleveland Clinic Foundation, Cleveland, Ohio, and SurVivaLink Corp, Minneapolis, Minn (J.E.B., K.F.O.).

Correspondence to Patrick J. Tchou, MD, Director, Clinical Cardiac Electrophysiology, Department of Cardiology/F15, Cleveland Clinic Foundation, 9500 Euclid Ave, Cleveland, OH 44195. E-mail tchoup{at}cesmtp.ccf.org


*    Abstract
up arrowTop
*Abstract
down arrowIntroduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
Background—Phase-2 voltage and maximum pulse width are dependent on phase-1 pulse characteristics in a single-capacitor biphasic waveform. The use of 2 separate output capacitors avoids these limitations and may allow waveforms with lower defibrillation thresholds. A previous report also suggested that the optimal tilt may be >70%. This study was designed to determine an optimal biphasic waveform by use of a combination of 2 separate and fully (95% tilt) discharging capacitors.

Methods and Results—We performed 2 external defibrillation studies in a pig ventricular fibrillation model. In group 1, 9 waveforms from a combination of 3 phase-1 capacitor values (30, 60, and 120 µF) and 3 phase-2 capacitor values (0=monophasic, 1/3, and 1.0 times the phase-1 capacitor) were tested. Biphasic waveforms with phase-2 capacitors of 1/3 times that of phase 1 provided the highest defibrillation efficacy (stored energy and voltage) compared with corresponding monophasic and biphasic waveforms with the same capacitors in both phases except for waveforms with a 30-µF phase-1 capacitor. In group 2, 10 biphasic waveforms from a combination of 2 phase-1 capacitor values (30 and 60 µF) and 5 phase-2 capacitor values (10, 20, 30, 40, and 50 µF) were tested. In this range, phase-2 capacitor size was more critical for the 30-µF phase-1 than for the 60-µF phase-1 capacitor. The optimal combinations of fully discharging capacitors for defibrillation were 60/20 and 60/30 µF.

Conclusions—Phase-2 capacitor size plays an important role in reducing defibrillation energy in biphasic waveforms when 2 separate and fully discharging capacitors are used.


Key Words: defibrillation • death, sudden • ventricles


*    Introduction
up arrowTop
up arrowAbstract
*Introduction
down arrowMethods
down arrowResults
down arrowDiscussion
down arrowReferences
 
The use of an exponential biphasic waveform has been shown to be superior to an exponential monophasic waveform in external1 as well as internal2 3 defibrillation. Use of a single capacitor for generating exponential biphasic waveforms poses several limitations, however. Increasing the phase-1 duration may not generate the optimal phase-2 leading-edge voltage or delivered charge, because phase-2 leading-edge voltage is dependent on phase-1 pulse width, phase-1 leading-edge voltage, capacitance, and impedance. Therefore, the optimal single-capacitor biphasic waveform may not generate the best combination of phase 1 and phase 2 compared with the optimal combination of 2 separate capacitors for the 2 phases.

In a previous report,4 we found that extending the phase-1 tilt of a biphasic waveform from 30% to 70% lowered the defibrillation threshold (DFT) energy when a 60-µF phase-1 capacitor was used. This finding suggested that the optical tilt may be >70%. Second, that previous report4 compared the use of a phase-2 capacitor (20 µF) that was smaller than the phase-1 capacitor with the use of same-size capacitors in both phases. The results indicated that the use of a smaller capacitor in phase 2, when charged to the same leading-edge voltage as phase 1, yielded the optimal DFT energies.

External defibrillation efficacy is influenced by multiple factors, such as electrode pad size,5 6 shock waveform,1 7 8 9 10 and sequential shocks.11 Because capacitor size may also play a role in determining external defibrillation efficacy,12 optimizing capacitor sizes may contribute to maximizing defibrillation efficacy. Theoretical models, assuming a 50-{Omega} impedance, have suggested that capacitors in the 30- to 60-µF range may provide the optimal capacitor for phase 1 of a biphasic waveform.13 14 Use of a smaller capacitor in a fully discharging capacitor waveform may be particularly important, because larger capacitors could result in exceedingly long pulse widths when the defibrillation shock is applied through a higher-impedance pathway. For example, for transthoracic impedance of 80 to 100 {Omega},8 9 10 15 discharging a 120-µF capacitor to 95% tilt would take >30 ms.

The purpose of this study was to assess the optimal capacitor sizes (phase 1 and phase 2) for a fully discharging (95% tilt) biphasic waveform in the above-discussed range of 30, 60, and 120 µF that could be implemented in a standard external defibrillation device.


*    Methods
up arrowTop
up arrowAbstract
up arrowIntroduction
*Methods
down arrowResults
down arrowDiscussion
down arrowReferences
 
The use of experimental animals in this study was approved by the Animal Research Committee of the Cleveland Clinic Foundation and conformed to the recommendations of the American Heart Association on Research Animal Use.

Study Preparation and Surgical Procedure
Anesthetized pigs were used in this study. Animal preparation and surgical procedures have been described in detail in previous publications.4 16 Briefly, the swine were intubated with a cuffed endotracheal tube and ventilated with room air supplemented with oxygen through a Drager SAV respirator (North American Drager), which was adjusted as needed to maintain normal arterial blood gases. A transvenous unipolar defibrillation lead (model 497, Intermedics Inc) was inserted into the right ventricular (RV) apex through the right external jugular vein under fluoroscopy. Three adhesive pad electrodes, each with a surface area of 75 cm2, were applied to the right upper pectoral, the left upper pectoral, and the cardiac apical area on shaved skin.

Defibrillation Protocol
The 3 adhesive pad electrodes and the RV apex defibrillation lead were connected to an external defibrillator custom-made by SurVivaLink Corp. The function of this external defibrillator was described in detail in our previous publication.4 Ventricular fibrillation (VF) was induced by delivery of 60-Hz AC (15 V) for 3 seconds through the RV apex defibrillation lead and the left pectoral pad electrode. After VF sustained for 10 seconds, 1 of the test waveforms was delivered between the right upper pectoral pad electrode and the apical pad electrode at the end-respiration phase. The apical pad electrode was the anode for phase 1 of the biphasic waveforms. If defibrillation failed for the test shock waveform, a rescue shock (450 to 900 V) was delivered between the RV defibrillation lead and the left upper pectoral pad electrode. A recovery period of >=3 minutes was allowed between each episode of VF. VF was not reinitiated until heart rate and blood pressure returned to the preshock values.

Defibrillation Waveforms
Two groups of experiments were performed for this report. In group 1, 3 phase-1 capacitor sizes (30, 60, and 120 µF) and 3 phase-2 capacitor sizes (0=monophasic, 1/3 of, and equal to phase-1 capacitor) were tested for a total of 9 different test shock waveforms. These 9 waveforms consisted of 3 exponential monophasic and 6 exponential biphasic waveforms, as illustrated in Figure 1ADown.



View larger version (17K):
[in this window]
[in a new window]
 
Figure 1. Diagram of each exponential waveform in group 1 and group 2. A, Top, 30-µF monophasic and 30/10- and 30/30-µF biphasic waveforms. Middle, 60-µF monophasic and 60/20- and 60/60-µF biphasic waveforms. Bottom, 120-µF monophasic and 120/40- and 120/120-µF biphasic waveforms. B, Top, 30/10-, 30/20-, 30/30-, 30/40-, and 30/50-µF biphasic waveforms. Bottom, 60/10-, 60/20-, 60/30-, 60/40-, and 60/50-µF biphasic waveforms. All waveforms have 95% tilt in phase 1 and phase 2. All biphasic waveforms use 2 separate capacitors, 1 for phase 1 and 1 for phase 2. Phase-2 leading-edge voltage is equal to phase-1 leading-edge voltage in all biphasic waveforms.

In group 2, a more detailed assessment of optimal phase-2 capacitor was performed using phase-1 capacitor sizes of 30 and 60 µF. Because of concerns regarding the potential length of phase-1 discharge when 120-µF capacitors were used and the experimental limitation of the number of waveforms that can be tested in a single experiment, the 120-µF phase-1 capacitor was not evaluated. Five phase-2 capacitor sizes (10, 20, 30, 40, and 50 µF) were used in combination with the 30- and 60-µF phase-1 capacitor. Ten different exponential biphasic waveforms were tested in this group to evaluate defibrillation efficacy, as shown in Figure 1BUp.

Evaluation of Defibrillation Efficacy
Defibrillation efficacy of each waveform was estimated by V50, defined as the leading-edge voltage of the waveform associated with a 50% likelihood of successful defibrillation. V50 was measured by a previously described Bayesian estimation technique.4 17 Ten defibrillation tests were performed for each waveform to evaluate V50. The first shock phase-1 leading-edge voltage was 1650 V in all waveforms. Sequential step changes in voltage were 350, 200, 150, 150, 100, 100, 50, 50, and 0 V. These steps in voltage change were either positive or negative, depending on failure or success in defibrillation of the preceding shock, respectively. E50 was the energy stored in the capacitors calculated from these V50 values.

In each animal, the effect of the first shock (1650 V) was tested first in a randomized order for all waveforms. Then, second shock voltages for the waveforms were determined, based on the result of the first, and were applied in random order as well. Thereafter, by the same procedure, the third through 10th shocks were delivered.

Statistical Analysis
Data in all DFT parameters were expressed as mean values±SD. Repeated-measures 1-way ANOVA was used to compare DFT parameters in group 1 and group 2. Pairwise comparisons of the waveforms were made for each parameter, with the least significant difference test used in group 1 and group 2. The null hypothesis was rejected for P<0.05.


*    Results
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
*Results
down arrowDiscussion
down arrowReferences
 
Group 1
A complete data set was obtained for 10 pigs (36±2 kg). All parameters at DFT are shown in Table 1Down.


View this table:
[in this window]
[in a new window]
 
Table 1. DFT Parameters and Waveform Characteristics in Group 1

Figure 2Down shows the E50 of stored energy and V50 of phase-1 leading-edge voltage for all waveforms. For waveforms using a 30-µF phase-1 capacitor, there was a trend toward lower E50 in the biphasic waveforms, but this difference did not reach statistical significance. However, there was a significant drop in leading-edge voltage for the biphasic waveforms, with the lowest voltage seen in the 30/30-µF waveform. In the waveforms using a 60-µF phase-1 capacitor, the optimal E50 and V50 were seen with use of the 20-µF phase-2 capacitor (P=0.008 and P=0.013, respectively). In the waveforms using a 120-µF phase-1 capacitor, the optimal E50 and V50 were observed with a phase-2 capacitor of 40 µF. Interestingly, using a phase-2 capacitor equal to phase 1 (120/120 µF) resulted in E50 and V50 that were significantly higher even compared with the corresponding monophasic (120-µF) waveform. Thus, the biphasic waveform with a smaller phase-2 capacitor than phase-1 capacitor reduced the E50 of stored energy when 60- and 120-µF capacitors were used in phase 1. For a phase-1 30-µF capacitor, the 30/30-µF waveform appears to be optimal when considering V50, even though the E50 estimates did not reach statistical differences. These results suggest that a phase-2 capacitor of 20 to 40 µF may be optimal for a wide range of phase-1 capacitors.



View larger version (27K):
[in this window]
[in a new window]
 
Figure 2. Mean and SD of E50 of stored energy and V50 of phase-1 leading-edge voltage for 3 monophasic and 6 biphasic waveforms in group 1. Left and right vertical axes show E50 of stored energy (J) and V50 of phase-1 leading-edge voltage (V), respectively; horizontal axis indicates each waveform.

Group 2
A complete data set was obtained for 10 pigs (33±4 kg). All parameters at DFT are shown in Table 2Down.


View this table:
[in this window]
[in a new window]
 
Table 2. DFT Parameters and Waveform Characteristics in Group 2

Figure 3Down shows E50 and V50 of the experimental results. In both phase-1 capacitor waveforms, the E50 and V50 values showed a minimum at phase-2 capacitor values of 20 or 30 µF. For a 30-µF phase-1 capacitor, the E50 and V50 rose markedly once phase-2 capacitor size exceeded 30 µF. For a 60-µF phase-1 capacitor, however, this rise was much less prominent. This difference may be related to the ratio of phase-1 to phase-2 capacitors and their effects on pulse width of the phases.



View larger version (25K):
[in this window]
[in a new window]
 
Figure 3. Mean and SD of E50 of stored energy and V50 of phase-1 leading-edge voltage for 10 biphasic waveforms in group 2. Left and right vertical axes show E50 of stored energy (J) and V50 of phase-1 leading-edge voltage (V), respectively; horizontal axis indicates each waveform.


*    Discussion
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
*Discussion
down arrowReferences
 
The results of this study show that DFT voltage and/or energy can be significantly lowered with the use of biphasic fully discharging capacitor waveforms compared with their corresponding monophasic waveforms. It is interesting to compare the effects of the phase-2 capacitor on the DFT. When a 30-µF phase-1 capacitor was used, the addition of a phase-2 capacitor (10 or 30 µF) lowered both DFT energy and voltage by {approx}40%. The lowest DFT was associated with the 30-µF phase-2 capacitor. With the 60- and 120-µF phase-1 capacitors, adding the corresponding 20- and 40-µF phase-2 capacitors had similar effects in lowering the DFT. However, when the corresponding phase-2 capacitors were made equal to the phase-1 capacitors, DFTs increased again, and for the 120-µF capacitors, biphasic thresholds were even higher than monophasic thresholds. These results would suggest that the optimal phase-2 capacitor for 30- to 120-µF phase-1 capacitors is in the range of 20 to 40 µF regardless of the capacitor used in phase 1. The results of these experiments showed that the capacitor combinations of 60/20 and 120/40 µF were clearly superior to the corresponding ones using the same capacitors for both phase 1 and phase 2 (60/60 and 120/120 µF). Perhaps because 30 µF was in the range of the optimal phase-2 capacitor, the difference between the 30/10- and the 30/30-µF waveforms was small.

The results shown in Table 1Up reveal that the mean impedances in our experimental model were in the 40- to 50-{Omega} range. However, typical transthoracic impedances for human defibrillation are higher, 60 to 80 {Omega}, with values in the 80- to 100-{Omega} range seen occasionally.8 9 10 The use of a 120-µF capacitor may be problematic in this impedance range. Based on the pulse widths shown in Table 1Up, one can extrapolate that the use of the same 120-µF waveforms in humans can extend the pulse widths to the 20- to 30-ms range. Such long pulse widths are likely to be not as effective for defibrillation, because their duration will be well beyond the chronaxie of the defibrillation strength-duration curves.14 Thus, group 2 data shown in Table 2Up were limited to a more detailed analysis of the 30- and 60-µF waveforms.

As illustrated in Figure 3Up, the pattern of changes in the DFT voltage and energy associated with changing phase-2 capacitors was different when the 30-µF and the 60-µF phase-1 capacitors were compared. For the 30-µF phase-1 capacitor, there was a clear minimum V50 associated with the use of a 30-µF capacitor for phase 2. This minimum was reflected in the E50 values as well. When the size of the phase-2 capacitor exceeded that of phase 1, the DFT values rose markedly. However, in the 60-µF phase-1 capacitor, the variation in the phase-2 capacitor did not have as clear a minimum as that shown for the 30-µF phase-1 data, perhaps partly because a phase-2 capacitor size that exceeded that of phase 1 was lacking. A minimum for both E50 and V50, however, was seen at a phase-2 capacitor value of {approx}20 to 30 µF. Like the observations in internal defibrillation,18 19 the results of this study showed that exponential biphasic waveforms with a very small capacitor (20 to 40 µF) in phase 2 were effective for external defibrillation. Furthermore, the 60-µF waveform had lower V50 values than the 30-µF waveforms. We would postulate that the 60-µF biphasic waveform with a phase-2 capacitor of {approx}20 to 30 µF may be the optimal combination among the ones reported here for human external defibrillation. The 120/40-µF waveform had a lower-voltage DFT at similar-energy DFT but would probably generate excessively long pulses for the higher human impedances.

Fully Discharging Capacitor Waveform
Because the tail of the defibrillation pulse may act to refibrillate the myocardium, an appropriate truncation of waveforms has been postulated to reduce DFT energy.13 However, previous publications20 21 have reported somewhat conflicting results of truncation in monophasic capacitor discharge waveforms. The effect of truncating phase 1 in biphasic waveforms has not been studied in detail for defibrillation with a dual-capacitor system. Our recent external defibrillation study4 showed that DFT energy decreased with phase-1 tilt, increasing from 30% to 70% when a separate phase-2 capacitor was used. This finding suggested that the optimal phase-1 tilt for dual-capacitor biphasic waveforms may be >70%. Despite previous experimental observations that truncation is important in monophasic waveforms, the data reported here raise the interesting possibility that such truncation may not play as significant a role in biphasic waveforms. The results with the 120-µF phase-1 waveform are an example of this possibility. Despite use of a full-tilt phase 1 that has relatively high E50 and V50 when tested as a monophasic waveform, the addition of a small phase-2 capacitor markedly decreased the DFT to values that were even better than those for the 30- and 60-µF capacitors.

Changing Capacitance at Phase Reversal
A recent external defibrillation study1 showed the superiority of single-capacitor biphasic waveforms over exponential monophasic waveforms. However, there are a few limitations in such a single-capacitor biphasic waveform. Maximizing the phase-1 pulse width may not generate the optimal phase-2 leading-edge voltage or charge transfer, because phase-2 leading-edge voltage is dependent on phase-1 pulse width. In addition, several recent defibrillation studies19 22 23 have shown that a biphasic waveform with a smaller phase-2 capacitor can achieve lower DFT energy when a phase-1 capacitor of 60 µF is used. Therefore, optimal biphasic waveforms may be best generated with 2 separate capacitors, 1 for phase 1 and 1 for phase 2.

As shown in Figure 2Up, the addition of an appropriate phase-2 capacitor can improve the E50 of the biphasic waveform over that obtained with the corresponding monophasic waveform. However, the relative improvement of the E50 appears to be greater when larger capacitors are used in phase 1. Specifically, the biphasic waveforms of 30/10, 60/20, and 120/40 µF reduced E50 by 38%, 67%, and 81%, respectively, compared with their respective monophasic waveforms. Thus, the beneficial effects of adding a phase-2 capacitor to the monophasic waveform appears to depend on the capacitor size in phase 1. Although the monophasic 120-µF waveform had the highest E50 of the 3 capacitors, the greater improvement in E50 seen with the addition of phase 2 made the 120/40-µF waveform perform as well as the 60/20-, 30/10-, and 30/30-µF waveforms. Another interesting observation is the effect of having the same capacitor/pulse width in both phases of the biphasic waveform. Although the 120/120-µF waveform had a poor E50 compared with the 120/40-µF waveform, the difference was much less prominent when the 60/60-µF and the 60/20-µF waveforms were compared, and the difference was nonexistent when the 30/30-µF and the 30/10-µF waveforms were compared. This observation would suggest that the optimal capacitor for a phase-2 fully discharging waveform may be fairly constant, {approx}20 to 40 µF, regardless of the phase-1 capacitor used. Because phase-2 capacitors were charged to the same voltage as phase-1 capacitors in these experiments, this observation would imply that the actual need for charge transfer in phase 2 may be less for higher-capacitor phase-1 waveforms where the leading-edge voltage is lower, even though the pulse width is longer, at least within the impedance range seen in our experimental preparation. Thus, the need for phase-2 charge transfer may be related to the leading-edge voltage of phase 1 regardless of the phase-1 capacitor.

Applicability of the 120-µF Capacitor to Human External Defibrillation
The impedance for external shock in the human chest is 60 to 80 {Omega} and may exceed 100 {Omega} in some cases.8 9 10 15 Conversely, the shock impedance of our pig model was {approx}40 to 50 {Omega} in this study. Although the 120/40-µF biphasic waveform performed quite well, as evidenced by the data presented in Table 1Up and Figure 2Up, its applicability to higher-impedance pathways may be problematic. The phase-1 pulse width of the 120-µF waveforms tested here was {approx}14 ms with 40-{Omega} impedance. For clinical circumstances, this phase-1 pulse width may become 20 to 30 ms because of the higher impedance and may be even higher in high-impedance cases. Such a long pulse width would most likely be inefficient for external defibrillation. Thus, our group 2 experiments did not include this capacitor.

Clinical Implications
To facilitate widespread dissemination and public use of such external defibrillators, it is important that these devices be reliable and appropriately priced. Thus, technological approaches that would minimize the cost of construction may be an important consideration in public access defibrillators.

Full-tilt waveforms have several potential advantages over traditional biphasic waveform designs. These advantages include higher reliability and simpler design. All of this translates into a potentially better defibrillator. In a single-capacitor biphasic waveform, a switching system is necessary to change polarity at a high voltage of phase reversal and for truncation of phase 2. Conventional biphasic waveforms have always been generated with an H-bridge–style construction. This method used electronic switches to reverse the capacitor during the middle of the discharge cycle. Such a reversal requires switches that are capable of controlling both high currents and high voltages. Switching these high currents necessitates the use of isolated-gate, bipolar transistors (IGBTs), which are fairly bulky and relatively expensive. Because the polarity of the capacitor must be totally reversed, 4 IGBTs are required. All 4 must be carefully sequenced to properly reverse the voltage. In addition, dump resistors that are frequently used to dispose of residual charges on a capacitor after truncation would not be necessary with a nearly full-tilt waveform.

A 2-capacitor construction to implement the waveforms reported here would simplify the switching requirements, because the polarity of the capacitors does not need to be reversed. Two separate capacitor banks would be used for the 2 phases. This waveform design uses the same leading-edge voltage in both phases, which allows the use of a single-charge transformer. Because virtually all of the energy is delivered to the patient, truncation circuitry is greatly simplified. This simplification means fewer parts and less energy switching, which improves the reliability of the system while reducing space and power requirements.

Conclusions
The major findings of this external defibrillation study are as follows. (1) For biphasic waveforms using 2 separate and almost completely discharging (95% tilt) capacitors, the phase-2 capacitor size is an important factor in maximizing defibrillation efficacy. (2) Biphasic waveforms using 2 separate and fully discharging capacitors appear to function best with a phase-2 capacitor {approx}20 to 40 µF for phase-1 capacitors in the 30- to 120-µF range. (3) In an external defibrillator for human use, the 60/20- or 60/30-µF may be the optimal choice among the waveforms tested here, because the 120-µF capacitor may generate exceedingly long pulse widths and the 30-µF model would use higher voltages, possibly generating greater costs without additional benefits. (4) The use of simpler components in a high-tilt biphasic waveform as described here may improve the reliability and the expense of manufacturing external defibrillators.


*    Acknowledgments
 
This work was supported by a grant from the SurVivaLink Corporation. We would like to thank Donald G. Hills for his technical assistance in the care of the animals before and during the experiments.

Received March 10, 1999; revision received April 28, 1999; accepted May 26, 1999.


*    References
up arrowTop
up arrowAbstract
up arrowIntroduction
up arrowMethods
up arrowResults
up arrowDiscussion
*References
 

  1. Gliner BE, Lyster TE, Dillion SM, Bardy GH. Transthoracic defibrillation of swine with monophasic and biphasic waveforms. Circulation. 1995;92:1634–1643.[Abstract/Free Full Text]
  2. Walcott GP, Walcott KT, Knisley SB, Zhou X, Ideker RE. Mechanisms of defibrillation for monophasic and biphasic waveforms. Pacing Clin Electrophysiol. 1994;17:478–498.[Medline] [Order article via Infotrieve]
  3. Blanchard SM, Ideker RE. Mechanisms of electrical defibrillation: impact of new experimental defibrillator waveforms. Am Heart J. 1994;127:970–977.[Medline] [Order article via Infotrieve]
  4. Yamanouchi Y, Brewer JE, Mowrey KA, Donohoo AM, Wilkoff BL, Tchou PJ. The optimal small-capacitor biphasic waveform for external defibrillation: influence of phase-1 tilt and phase-2 voltage. Circulation. 1998;98:2487–2493.[Abstract/Free Full Text]
  5. Hoyt R, Grayzel J, Kerber RE. Determinants of intracardiac current in defibrillation: experimental studies in dogs. Circulation. 1981;64:818–823.[Free Full Text]
  6. Dalzell GW, Cunningham SR, Anderson J, Adgey AA. Electrode pad size, transthoracic impedance and success of external ventricular defibrillation. Am J Cardiol. 1989;64:741–744.[Medline] [Order article via Infotrieve]
  7. Wilson CM, Bailey A, Allen JD, Anderson J, Adgey AA. Transthoracic defibrillation threshold of sine and trapezoidal waveforms in defibrillation. J Electrocardiol. 1989;22:241–247.[Medline] [Order article via Infotrieve]
  8. Greene HL, DiMarco JP, Kudenchuk PJ, Scheinman MM, Tang AS, Reiter MJ, Echt DS, Chapman PD, Jazayeri MR, Chapman FW, Ahmed M, Johnson JL, Niskanen RA, Biphasic Waveform Defibrillation Investigators. Comparison of monophasic and biphasic defibrillating pulse waveforms for transthoracic cardioversion. Am J Cardiol. 1995;75:1135–1139.[Medline] [Order article via Infotrieve]
  9. Bardy GH, Gliner BE, Kudenchuk PJ, Poole JE, Dolack GL, Jones GK, Anderson J, Troutman C, Johnson G. Truncated biphasic pulses for transthoracic defibrillation. Circulation. 1995;91:1768–1774.[Abstract/Free Full Text]
  10. Bardy GH, Marchlinski FE, Sharma AD, Worley SJ, Luceri RM, Yee R, Halperin BD, Fellows CL, Ahern TS, Chilson DA, Packer DL, Wilber DJ, Mattioni TA, Reddy R, Kronmal RA, Lazzara R. Multicenter comparison of truncated biphasic shocks and standard damped sine wave monophasic shocks for transthoracic ventricular defibrillation. Circulation. 1996;94:2507–2514.[Abstract/Free Full Text]
  11. Kerber RE, Spencer KT, Kallok MJ, Birkett C, Smith R, Yoerger D, Kieso RA. Overlapping sequential pulses: a new waveform for transthoracic defibrillation. Circulation. 1994;89:2369–2379.[Abstract/Free Full Text]
  12. Peleska B. Optimal parameters of electrical impulses for defibrillation by condenser discharges. Circ Res. 1966;18:10–17.[Abstract/Free Full Text]
  13. Irnich W. Optimal truncation of defibrillation pulses. Pacing Clin Electrophysiol. 1995;18:673–688.
  14. Kroll MW. A minimal model of the monophasic defibrillation pulse. Pacing Clin Electrophysiol. 1993;16:769–777.[Medline] [Order article via Infotrieve]
  15. Kerber RE, Jensen SR, Gascho JA, Grayzel J, Hoyt R, Kennedy J. Determinants of defibrillation: prospective analysis of 183 patients. Am J Cardiol. 1983;52:739–745.[Medline] [Order article via Infotrieve]
  16. Yamanouchi Y, Mowrey KA, Niebauer MJ, Tchou PJ, Wilkoff BL. An additional lead improves defibrillation efficacy with an abdominal "hot can" electrode system. Circulation. 1997;96:4400–4407.[Abstract/Free Full Text]
  17. Compos AT, Malkin RA, Ideker RE. An up-down Bayesian, defibrillation efficacy estimator. Pacing Clin Electrophysiol. 1997;20:1292–1300.[Medline] [Order article via Infotrieve]
  18. Walcott GP, Rollins DL, Smith WM, Ideker RE. Effect of changing capacitors between phases of a biphasic defibrillation shock. Pacing Clin Electrophysiol. 1996;19:945–954.[Medline] [Order article via Infotrieve]
  19. Yamanouchi Y, Mowrey KA, Kroll MW, Brewer JE, Donohoo AM, Wilkoff BL, Tchou PJ. Optimized first phase tilt in "parallel-series" biphasic waveform. J Cardiovasc Electrophysiol. 1997;8:649–657.[Medline] [Order article via Infotrieve]
  20. Walcott GP, Walker RG, Cates AW, Krassowska W, Smith WM, Ideker RE. Choosing the optimal monophasic and biphasic waveforms for ventricular defibrillation. J Cardiovasc Electrophysiol. 1995;6:737–750.[Medline] [Order article via Infotrieve]
  21. Gold MR, Shorofsky SR. Strength-duration relationship for human transvenous defibrillation. Circulation. 1997;96:3517–3520.[Abstract/Free Full Text]
  22. Yamanouchi Y, Mowrey KA, Nadzam GR, Hills DG, Kroll MW, Brewer JE, Donohoo AM, Wilkoff BL, Tchou PJ. Large change in voltage at phase reversal improves biphasic defibrillation thresholds: parallel-series mode switching. Circulation. 1996;94:1768–1773.[Abstract/Free Full Text]
  23. Yamanouchi Y, Brewer JE, Hills DG, Donohoo AM, Steinkogler GL, Tchou PJ. The external biphasic waveform with changing capacitor at phase reversal reduced defibrillation energy. Circulation. 1997;96(suppl I):I-530. Abstract.



This article has been cited by other articles:


Home page
JAMAHome page
J. P. Marenco, P. J. Wang, M. S. Link, M. K. Homoud, and N. A. M. Estes III
Improving Survival From Sudden Cardiac Arrest: The Role of the Automated External Defibrillator
JAMA, March 7, 2001; 285(9): 1193 - 1200.
[Abstract] [Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yamanouchi, Y.
Right arrow Articles by Tchou, P. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yamanouchi, Y.
Right arrow Articles by Tchou, P. J.
Related Collections
Right arrow CPR and emergency cardiac care
Right arrow Arrhythmias, clinical electrophysiology, drugs